Abstract

Cryoablation is a minimally invasive ablation technique for primary and metastatic hepatic tumors. Inadequate freezing around large blood vessels due to the warm blood flow can lead to local recurrence, and thus, necessitates close application of a cryoprobe to the large blood vessels. In this study, we constructed a perfusion model with an ex vivo bovine liver and ablated the tissue around a large blood vessel with one or two cryoprobes applied to the side of the vessel. The finite-element computer model developed in our previous study was modified to include a blood vessel and its convective heat transfer to the vicinity of the blood vessel. We compared the predicted simulation results to those acquired from this ex vivo perfusion model. The results indicate that blood vessels act as a heat source and generate steep temperature profiles in the area next to the large blood vessel. After validation, the maximum allowable distance between the cryoprobe and the large blood vessel for successful cryoablation was presented. The results of this study should be considered when placing cryoprobes in the vicinity of large blood vessels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.