Abstract

Carbon fiber composites are emerging as a promising new biomaterial for chest wall reconstruction implants due to their mechanical properties and biocompatibility. This work evaluates the biomechanics of carbon fiber artificial ribs using finite element analysis and clinical implementation. Static simulations of normal breathing process show the maximum stress on the implant is only 2.83 MPa, far below the material ultimate strength of 60 MPa, indicating the excellent fit for maintaining respiratory function. Dynamic collision simulations demonstrate the artificial rib model could withstand a 4 kg rigid object impact at 2 m/s without fracture. Reconstructing the artificial rib with a human rib in the finite element analysis model increases the overall stress tolerance. The impact force required for fracture increases 48% compared to the artificial rib alone, suggesting improved strength from rib integration. Clinically, 10 of 13 patients receiving the artificial rib implants show no significant loss of pulmonary function based on spirometry tests. Based on our findings, the combined simulations and clinical results validate the strong mechanical performance and biocompatibility of the carbon fiber artificial ribs for chest wall reconstruction under static and dynamic loading while maintaining normal respiratory function.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call