Abstract

A fingerprint appearance to the macula was observed in both eyes of a patient with foveal hypoplasia and pendular horizontal nystagmus imaged with the Optos California (Marlborough, MA) fundus camera. (A, B) Examination under anesthesia with Natus Retcam 3 (San Carlo, CA) color fundus photography (C) and Bioptigen Envisu C2300 (Buffalo Grove, IL) portable spectral-domain OCT (D) confirmed these to be artifactual. The Optos camera utilizes sweeping red and green lasers and multi-element optics with nonlinear surfaces resulting in nonuniform scan velocities slowest across the fovea.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.