Abstract

Volumetric absorptive microsampling (VAMS) is increasingly proposed as a clinically reliable therapeutic drug monitoring (TDM) sampling methodology. The study aimed to establish the reliability and real-life feasibility of patient self-collected capillary VAMS for TDM of antiseizure medication (ASMs), using plasma ASMs concentrations from venous blood as a reference standard. Nurses collected venous and capillary blood samples using VAMS. Afterward, persons with epilepsy (PWE) performed VAMS sampling by themselves. All samples were analyzed by UHPLC-MS/MS. We performed a cross-validation study, comparing ASMs concentrations obtained by VAMS nurses and patients’ self-collected versus plasma through Bland-Altman analysis and Passing-Bablok regression. We enrolled 301 PWE (M: F 42.5%:57.5%; mean age 44±16 years), treated with 13 ASMs, providing a total of 464 measurements. Statistical analysis comparing VAMS self-collected versus plasma ASMs concentrations showed a bias close to zero and slope and intercept values indicating a good agreement for CBZ, LCS, LEV, LTG, OXC, PB, and PHT, while a systematic difference between the two methods was found for VPA, PMP, TPM and ZNS. This is the first study showing the reliability and feasibility of the real-world application of PWE self-collected VAMS for most of the ASMs considered, giving a promising basis for at-home VAMS applications.