Abstract

This report describes a comparison of 10 pF and 100 pF capacitance standards organized within theframework of EUROMET Project No. 345 between 1995 and 1998. Piloted by the National PhysicalLaboratory (NPL, UK), the comparison involved sixteen national metrology institutes (NMIs), several ofwhich are members of other regional metrology organizations, and the Bureau International des Poids etMesures. The results show that there are significant differences between representations of the farad by some NMIs. However, the agreement demonstrated by the comparison gives confidence in maintaining traceability for the farad via either a calculable capacitor or a quantum Hall reference standard and the consensus value of the von Klitzing constant, RK-90.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.