Abstract
A regional key comparison APMP.AUV.A-K1 has been carried out for the pressure sensitivity of laboratory standard microphones. The NationalMetrology Institute of Japan (NMIJ) piloted this project. Two LS1P microphones were circulated through nine national metrology institutes and calibrated in the frequency range from 63 Hz to 8 kHz. Deviations from the mean value are below ± 0.05 dB for all the frequencies. Results have been linked to CIPM key comparison CCAUV.A-K1 by introducing a correction factor. For every participant, the degree of equivalence with a key comparison reference value is within the corresponding expanded uncertainty.Main text.To reach the main text of this paper, click on Final Report.The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.