Abstract

This is the final report of key comparison EURAMET.T-K7.1, a bilateral comparison of water triple point cells between the Czech Metrology Institute, CMI, and the Slovensky Metrologicky Ustav, SMU, of Slovakia. The comparison was requested by CMI, and the measurement protocol was compatible with the one used for key comparison EUROMET.T-K7. The results show a satisfactory degree of equivalence between SMU and CMI, and allow linkage to the parent key comparison EUROMET.T-K7. As a consequence, CMI is able to support CMCs at the temperature of the triple point of water and submit them for publication in the KCDB.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).