Abstract

A bilateral comparison has been organized between VSL, The Netherlands, and BIM, Bulgaria, of the realisations of the international temperature scale ITS-90 at the fixed points of Hg, H2O, Ga, In, Sn and Zn using a long-stem SPRT of very good stability as the transfer device. This comparison is registered as EURAMET project T-K3.1 in the BIPM key comparison database and its results are linked to those of key comparison CCT-K3.This comparison was organized in the framework of Phare project BG 2005/017-353.02.02, Lot 1, and is in this framework financed by the EU. This project ran from March 2008 to the end of February 2009.For all points of the measurements, a good agreement between the results obtained by BIM and in CCT-K3 could be demonstrated.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.