Abstract

The supplementary comparison COOMET.AUV.A-S1 for secondary calibration methods using WS1 and WS2 measurement microphones was carried out from 2009 to 2010. The results were submitted to and approved by CCAUV in April 2014.Four National Metrology Institutes took part in this comparison and are as follows: BelGIM (Belarus), VNIIFTRI (Russia), SMU (Slovakia) and DP NDI 'Sistema' (Ukraine). Three of the above NMIs (VNIIFTRI, SMU and DP NDI 'Sistema') had earlier participated in COOMET key comparisons and one NMI (VNIIFTRI) had also participated in CCAUV key comparisons.The Comparison Reference Values were calculated as the weighted mean values from results obtained by three institutes. The comparison results show agreement for all participants in the frequency range from 20 Hz to 12.5 kHz for WS1 microphones, and in the frequency range from 20 Hz to 16 kHz for WS2 microphones.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).