Abstract

A key comparison in dew-point temperature was carried out among the national standards held by NPL (pilot), NMIJ, INTA, VSL, INRIM, MIKES, NIST, NIM, VNIIFTRI-ESB and NMC. A pair of condensation-principle dew-point hygrometers was circulated and used to compare the local realisations of dew point for participant humidity generators in the range −50 °C to +20 °C. The duration of the comparison was prolonged by numerous problems with the hygrometers, requiring some repairs, and several additional check measurements by the pilot. Despite the problems and the extended timescale, the comparison was effective in providing evidence of equivalence. Agreement with the key comparison reference value was achieved in the majority of cases, and bilateral degrees of equivalence are also reported.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call