Abstract

This report describes the results obtained in a bilateral comparison of primary vibration calibration facilities. Two piezoelectric standard accelerometers (one single-ended as well one back-to-back type) were calibrated first at METAS, then at the PTB and again at METAS after their return.The METAS and the PTB calibrations were found to be in excellent agreement, with a mean difference smaller than the expanded uncertainty (k = 2) of the comparison.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.