Abstract
In order to assess the performance of the standards and techniques used for calibration and measurement of UVA irradiance responsivity of photodetectors in NMISA, South Africa, a new comparison was decided as a follow-up to comparison APMP.PR-S1. It is registered in the Key Comparison Data Base (KCDB) of BIPM as a bilateral supplementary comparison, with the identifier APMP.PR-S1.1. The comparison was carried out following the same technical protocol as that of supplementary comparison APMP PR-S1.The principle, organization and method of the comparison, as well as the preliminary measurements at the pilot laboratory NMC-A*STAR Singapore, were described in the Final Report of the APMP.PR-S1 comparison. The results of this bilateral comparison show that the NMISA's results lie within ±2% against the comparison reference values of APMP.PR-S1, which is a great improvement.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.