An international intercomparison project was performed to test the reproducibility and the comparative accuracy of the various absolute intensity calibration techniques in current use in small-angle X-ray scattering with the participation of fifteen investigators from eight different laboratories in six countries. In the project, the absolute differential X-ray scattering cross sections of standard samples of glassy carbon and polystyrene were calibrated using five different calibration techniques and two different X-ray wavelengths. The results have been intercompared with a variety of statistical techniques. It is concluded that angularly dependent errors associated with determining the zero of angle, dead-time corrections, collimation corrections, and insufficiently close data point spacing are more important in accounting for discrepancies between laboratories than are differences in the absolute intensity calibration methods themselves.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.