Abstract

Main textThe key comparison CCQM-K157 for the thickness measurement of HfO2 films was performed by the Surface Analysis Working Group (SAWG) of the Consultative Committee for Amount of Substance (CCQM). The aim of CCQM-K157 is to establish the measurement traceability and to ensure the equivalency in the measurement capability of national metrology institutes for the thickness measurement of HfO2 films.In this key comparison, the thicknesses of six HfO2 films with the nominal thickness range from 0.7 nm to 6 nm were compared by x-ray photoelectron spectroscopy (XPS), x-ray reflectometry (XRR), transmission electron microscopy (TEM), spectroscopic ellipsometry (SE) and medium energy ion scattering spectrometry (MEIS).To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call