Abstract

BackgroundRespiratory insufficiency is the major cause of mortality in patients with amyotrophic lateral sclerosis or Lou Gehrig’s disease. This is the final report of the diaphragm pacing (DP) pilot trial. MethodsPatients underwent laparoscopic diaphragm electrode implantations and subsequent conditioning of diaphragms. Serial respiratory function tests were performed in the initial year and followed until death. ResultsSixteen patients were implanted with no perioperative or unanticipated device-related adverse events. There were 452 implant-months of follow-up. DP allowed greater movement of the diaphragm under fluoroscopy, increased muscle thickness, and decreased the decline in forced vital capacity. Median survival from implant was 19.7 months with the cause of death respiratory in only 31%. ConclusionsLong-term analysis of DP in amyotrophic lateral sclerosis showed no safety issues and can positively influence diaphragm physiology and survival. This formed the initial basis for subsequent US Food and Drug Administration approval.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.