Abstract

In a prospective investigation the diagnostic accuracy of filmscreen and digital radiography in rheumatoid arthritis of hands was compared. Seventy hands of 36 patients with established rheumatoid arthritis were included in the study. Each of 11 joints in every hand was evaluated regarding the following radiologic parameters: soft tissue swelling, joint space narrowing, erosions and periarticular osteopenia. The digital images were obtained with storage phosphor image plates and evaluated in 2 forms; as digital hard-copy on film and on a monitor of an interactive workstation. The digital images had a resolution of either 3.33 or 5.0 lp/mm. ROC curves were constructed and comparing the area under the curves no significant difference was found between the 3 different imaging forms in either resolution group for soft tissue swelling, joint space narrowing and erosions. The film-screen image evaluation of periarticular osteopenia was significantly better than the digital hard-copy one in the 3.33 lp/mm resolu...