Abstract

The properties of aerosols generated from salbutamol sulfate solution (1 mg/mL) using an air-jet nebulizer were evaluated using Next Generation Impactor (NGI), a full cascade impactor, and Fast Screening Impactor (FSI), an abbreviated impactor measurement (AIM). Both impactors were operated under the same experimental conditions. The samples were recovered and assayed using validated high performance liquid chromatography (HPLC). The study investigated AIM-Human Respiratory Tract (HRT) concept by comparing key parameters of aerosolization i.e. fine particle dose (FPD) and fine particle fraction (FPF) measured using FSI, with NGI as baseline. The results showed that FSI yielded different but comparable values for FPD and FPF, indicating that it is alternative impactor to NGI. Despite the fact that FSI could not replace NGI, it may be used as an alternative impactor for simple and rapid aerosol characterization of formulations in some pharmaceutical development and quality control processes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.