Abstract

Here we describe baseline validation studies and field performance of a research-use-only chemiluminescent multiplex serology panel for measles, mumps, rubella, and varicella-zoster virus used with dried blood spots in support of the 2013-2014 Democratic Republic of the Congo Demographic and Health Survey. Characterization of the panel using U.S. FDA-cleared commercial kits shows good concordance for measles, mumps, rubella, and varicella-zoster with average sensitivity across assays of 94.9% and an average specificity of 91.4%. As expected, performance versus available standards validated for plaque-reduction assays does not provide a 1:1 correspondence with international units and yet demonstrates excellent linearity (average Hill's slope = 1.02) and ∼4 logs of dynamic range. In addition, for the four assays, the multiplexed format allowed for inclusion of three positive and two negative controls for each sample. A prototype Dynex Multiplier chemiluminescent automated immunoassay instrument with a charge-coupled device camera provided a rugged and robust processing and data acquisition platform. Performance of a multiplex instrument for serological testing in a substantially resource-limited environment shows excellent reproducibility, minimal cross-reactivity, and a clear discrimination between specific assays and should be considered a viable option for future serosurveys.IMPORTANCE The critical evaluation of immunization programs is key to identifying areas of suboptimal vaccination coverage, monitoring activities, and aiding development of public health policy. For evaluation of vaccine effectiveness, direct antibody binding assay methods, including enzyme immunoassay, enzyme-linked fluorescence assays, and indirect immunofluorescence assay, are most commonly used for detection of IgG antibodies. However, despite their well-demonstrated, reliable performance, they can be labor-intensive and time-consuming and require separate assays for each individual marker. This necessitates increased sample volumes, processing time, and personnel, which may limit assessment to a few key targets in resource-limited settings, that is, low- and middle-income locations where funding for public health or general infrastructure that directly impacts public health is restricted, limiting access to equipment, infrastructure, and trained personnel. One solution is a multiplexed immunoassay, which allows for the detection of multiple analytes in a single reaction for increased efficiency and rapid surveillance of infectious diseases in limited-resource settings. Thus, the scope of the project precluded a full validation, and here we present abbreviated validation studies demonstrating adequate sensitivity, specificity, and reproducibility.

Highlights

  • We describe baseline validation studies and field performance of a research-use-only chemiluminescent multiplex serology panel for measles, mumps, rubella, and varicella-zoster virus used with dried blood spots in support of the 2013–2014 Democratic Republic of the Congo Demographic and Health Survey

  • While the incidence of vaccine-preventable diseases (VPDs) such as measles, mumps, rubella, and varicella-zoster virus (VZV) infection (MMRV) has been substantially reduced in developed countries due to effective immunization programs, much work remains in low- and middle-income countries (LMICs) [1, 2]

  • While most countries have a national immunization program which provides a number of standard routine vaccinations for children and pregnant women, not all countries have introduced vaccines against MMRV into the national immunization framework, and administrative data may not be reliable for determining accurate rates of seroprotection [6]

Read more

Summary

Introduction

We describe baseline validation studies and field performance of a research-use-only chemiluminescent multiplex serology panel for measles, mumps, rubella, and varicella-zoster virus used with dried blood spots in support of the 2013–2014 Democratic Republic of the Congo Demographic and Health Survey. For evaluation of vaccine effectiveness, direct antibody binding assay methods, including enzyme immunoassay, enzyme-linked fluorescence assays, and indirect immunofluorescence assay, are most commonly used for detection of IgG antibodies Despite their well-demonstrated, reliable performance, they can be labor-intensive and time-consuming and require separate assays for each individual marker. ELISA is considered the gold standard for measuring soluble biomarkers and is widely accepted in clinical practice [15] Despite their well-demonstrated, reliable performance, ELISAs can be labor-intensive and time-consuming and require separate assays for each individual marker. This necessitates increased sample volumes, processing time, and personnel, which may limit assessment to a few key targets in resource-limited settings. During large-scale surveillance or screening activities, the collection of DBSs in lieu of wet specimens is common practice in LMICs given that DBS collection is minimally invasive, reduces cost for sample collection, requires less cold-chain management, and may be better accepted by the donor and/or parent [23]

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.