Abstract

We describe a membrane-based collection/analysis system that differentially monitors H2S and CH3SH, and to which a conductometric SO2 analyzer using the same collector was coupled. A diffusion scrubber (DS) comprised of a Nafion tube collects H2S selectively while a porous polytetrafluoroethylene (pPTFE) DS collects both H2S and CH3SH. Both gases are measured via their ability to react with fluorescein mercuric acetate (FMA) which results in decreased fluorescence. The limited dynamic range of a negative signal procedure was overcome by using dual DS units comprised of short and long scrubbers, placed serially in the liquid flow line. Different DS designs and membrane materials were investigated. H2S, CH3SH, and SO2 from a biogenic point source were continuously measured, and the H2S/CH3SH data compared well with a standard procedure involving Tedlar bag collection, preconcentration and thermal desorption from a Tenax trap, and measurement by gas chromatography/flame photometric detection. Walkaround portability of the instrument and very large dynamic range measurement of H2S and SO2 were demonstrated around the Mt. Aso volcano.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.