Abstract

The stability of the main magnetic field is critical for prolonged in vivo magnetic resonance spectroscopy (MRS) acquisitions, especially for difference spectroscopy. This study was focused on the implementation and optimization of a field-frequency lock (FFL) on a whole body spectrometer, to correct the main field drift during localized proton MRS of the human brain. The FFL was achieved through a negative feed-back applied in real time on the Z0 shim coil current, after calculation of the frequency shift from a reference signal. This signal was obtained from the whole head with a small flip angle acquisition interleaved with the PRESS acquisition of interest. To avoid propagation of the important short-term time-correlated fluctuations of the head water frequency (mainly due to respiratory motion) onto Z0 correction, the sampling rate of the reference frequency and the smoothing window for the Z0 correction were carefully optimized. Thus, an effective FFL was demonstrated in vivo with no significant increase of the short-term variance of the water frequency. Magn Reson Med 1999 42:636-642, 1999.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.