Abstract
Flatness and symmetry are main parameters determining the quality of a photon beam produced by linear accelerators. The quality assurance in routine clinical practice of radiotherapy and consequently the treatment-outcome depend definitely on the physical parameters of treatment-delivery. Several recommendations from national and international associations are reported. By reviewing the current literature and mainly according to the World Health Organization (WHO) report of quality assurance in radiotherapy, we may suggest that for flatness and symmetry, the optimal level of deviation should be within +/- 3%. Flatness and symmetry should be checked monthly or once a year in accordance to the guidelines of national societies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.