Abstract
BackgroundEarly infant HIV diagnosis (EID) coverage and uptake remains challenging. Point-of-care (POC) testing may improve access and turn-around-times, but, while several POC technologies are in development there are few data on their implementation in the field.MethodsWe conducted an implementation study of the Alere q Detect POC system for EID at two public sector health facilities in Cape Town. HIV-exposed neonates undergoing routine EID testing at a large maternity hospital and a primary care clinic received both laboratory-based HIV PCR testing per local protocols and a POC test. We analysed the performance of POC versus laboratory testing, and conducted semi-structured interviews with providers to assess acceptability and implementation issues.ResultsOverall 478 specimens were taken: 311 tests were performed at the obstetric hospital (median child age, 1 days) and 167 six-week tests in primary care (median child age, 42 days). 9.0% of all tests resulted in an error with no differences by site; most errors resolved with retesting. POC was more sensitive (100%; lower 95% CI, 39.8%) and specific (100%, lower 95% CI, 98%) among older children tested in primary care compared with birth testing in hospital (90.0%, 95% CI, 55.5–99.8% and 100.0%, lower 95% CI, 98.4%, respectively). Negative predictive value was high (>99%) at both sites. In interviews, providers felt the device was simple to use and facilitated decision-making in the management of infants. However, many wanted clarity on the cause of errors on the POC device to help guide repeat testing.ConclusionsPOC EID testing performs well in field implementation in health care facilities and appears highly acceptable to health care providers.
Highlights
Every year almost 1.5 million infants are exposed to HIV
Overall 478 specimens were taken: 311 tests were performed at the obstetric hospital and 167 six-week tests in primary care. 9.0% of all tests resulted in an error with no differences by site; most errors resolved with retesting
POC was more sensitive (100%; lower 95% confidence intervals (CI), 39.8%) and specific (100%, lower 95% CI, 98%) among older children tested in primary care compared with birth testing in hospital (90.0%, 95% CI, 55.5–99.8% and 100.0%, lower 95% CI, 98.4%, respectively)
Summary
We conducted an implementation study of the Alere q Detect POC system for EID at two public sector health facilities in Cape Town. HIV-exposed neonates undergoing routine EID testing at a large maternity hospital and a primary care clinic received both laboratory-based HIV PCR testing per local protocols and a POC test. We analysed the performance of POC versus laboratory testing, and conducted semi-structured interviews with providers to assess acceptability and implementation issues
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.