Abstract
We developed a field screening method for rapid analysis of Hg in water, soil, and sediment, which can be applied cost-effectively at Hg-contaminated sites. Samples are chemically pretreated in ordinary containers, followed by analysis of the sample headspace Hg vapor using a portable commercial analyzer. Hg in water samples is reduced directly by the addition of stannous chloride, while solids are first digested with aqua regia or piranha solution to liberate the Hg from the solids. Aided by vigorous agitation after adding the reductant, the elemental Hg partitions between solution and headspace according to Henry`s Law. The method requires about 2 and 15 minutes to complete for water and solids, respectively. The method provides very useful detection limits for water (0.1 {mu}g/L) and solids (2-3{mu}g/g). Intercomparisons with laboratory-analyzed environmental samples show good agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.