Abstract

IntroductionWith the introduction of accelerated partial breast irradiation (APBI) and the trend of reducing the number of fractions, the geometric accuracy of treatment delivery becomes critical. APBI patient setup is often based on fiducials, as the seroma is frequently not visible on pretreatment imaging. We assessed the motion of fiducials relative to the tumor bed between planning CT and treatment, and calculated margins to compensate for this motion. MethodsA cohort of seventy patients treated with APBI on a Cyberknife was included. Planning and in-room pretreatment CT scans were registered on the tumor bed. Residual motion of the centers of mass of surgical clips and interstitial gold markers was calculated. We calculated the margins required per desired percentage of patients with 100% CTV coverage, and the systematic and random errors for fiducial motion. ResultsFor a single fraction treatment, a margin of 1.8 mm would ensure 100% CTV coverage in 90% of patients when using surgical clips for patient set-up. When using interstitial markers, the margin should be 2.2 mm. The systematic and random errors were 0.46 mm for surgical clip motion and 0.60 mm for interstitial marker motion. No clinical factors were found predictive for fiducial motion. ConclusionsFiducial motion relative to the tumor bed between planning CT and APBI treatment is non-negligible and should be included in the PTV margin calculation to prevent geographical miss. Systematic and random errors of fiducial motion were combined with other geometric uncertainties to calculate comprehensive PTV margins for different treatment techniques.

Highlights

  • With the introduction of accelerated partial breast irradiation (APBI) and the trend of reducing the number of fractions, the geometric accuracy of treatment delivery becomes critical

  • As the seroma is frequently not visible on patient or target setup imaging and the fiducials are used as a surrogate of target localization, this motion should be included in the calculation of the planning target volume (PTV) margins to reduce the risk of geographical miss

  • To generalize our results to fractionated treatment schedules, we propose a value of 0.46 mm for the systematic and random error for positioning based on surgical clips

Read more

Summary

Introduction

With the introduction of accelerated partial breast irradiation (APBI) and the trend of reducing the number of fractions, the geometric accuracy of treatment delivery becomes critical. We assessed the motion of fiducials relative to the tumor bed between planning CT and treatment, and calculated margins to compensate for this motion. Residual motion of the centers of mass of surgical clips and interstitial gold markers was calculated. We calculated the margins required per desired percentage of patients with 100% CTV coverage, and the systematic and random errors for fiducial motion. The systematic and random errors were 0.46 mm for surgical clip motion and 0.60 mm for interstitial marker motion. Conclusions: Fiducial motion relative to the tumor bed between planning CT and APBI treatment is nonnegligible and should be included in the PTV margin calculation to prevent geographical miss. Systematic and random errors of fiducial motion were combined with other geometric uncertainties to calculate comprehensive PTV margins for different treatment techniques.

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.