Abstract

Extraction of passive fixation coronary sinus (CS) leads is typically easily achieved with manual traction. The ability to readily extract active fixation leads from the CS is less clear. Our first extraction experience with an active fixation CS lead was in a 58-year-old man with a 13-month-old Medtronic 4195 lead (Medtronic Inc., Minneapolis, MN, USA). The lobes of the lead would not fully undeploy. Significant, prolonged manual traction was required to free the lead from the cardiac vein. Inspection demonstrated fibrotic tissue growth into the lead lobes. Such growth may lead to an increase in extraction complications and failures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.