Abstract

A fiber-optic dipping liquid analyzer (FDLA) is developed for measuring liquid properties such as concentration, refractive index, surface tension, and viscosity. An important feature of the FDLA is that a liquid drop is introduced on the end face of a fiber probe, and the drop can be regarded as a planar-convex lens. The light transmitting path and receiving power are affected by the refractive index of the liquid drop. We present a theoretical and experimental analysis of the light transmission. A mathematical model of receiving power is established based on paraxial refraction imaging and fiber reflective intensity modulation methods. Sucrose-water solutions were tested with the FDLA. The experimental results agree well with the theoretical analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.