Abstract

The reliability of blood glucose monitoring in neonatology is not always confirmed. The aim of this study was to evaluate the reliability of blood glucose measurements made with three different devices in newborns. The study was prospective, conducted in a medical and neonatal intensive care department over a period of 4 months. Capillary glucose level was measured with three different glucometers and compared with venous glucose level determined using the hexokinase method. An ANOVA and Scheffe test were used for the correlation analysis. Three hundred and nine infants were included, with a mean age of 55h and a mean term of 39weeks of gestation. Mean blood glucose in the laboratory was 0.62±0.15g/L, 0.71±0.17g/L for Accu-Chek(®) Active, 0.80±0.17g/L for Accu-Chek(®) Performa, and 0.83±0.12g/L for Bionime. An ANOVA showed statistically significant differences between the measurements made by glucometers compared to the reference blood glucose levels, and the Scheffé method showed that glucometers overestimated the real plasma glucose levels. None of the devices used in this study was satisfactory. However, an estimation of blood glucose taking into consideration this numerical overestimation would allow early detection of hypoglycemia.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.