Abstract
A pH-sensitive detector for flow-through potentiometry based on a graphite/quinhydrone composite electrode was applied for flow-injection analysis (FIA) titrations. Hydrochloric acid and acetic acid were titrated by injection of samples into a sodium hydroxide carrier solution. System conditions were optimised by variation of flow rate, injection volume and titrant concentration. The parameters sampling frequency, residence time and dispersion coefficient were determined. The evaluation of peak width (time between the two inflection points on each side of the peak), peak area and slope of the raising edge of the peak lead were studied with respect to their use for calibration. Hydrochloric acid and acetic acid could be titrated down to a concentration of 2 x 10(-4) mol L(-1) using 150-microL injection volumes, which is almost ten times lower than can be achieved using colour indicators and a spectrophotometric detection.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.