Abstract

We developed a technique for direct start of peritoneal dialysis. Using a coiled or straight Tenckhoff catheter often results in obstruction of flow. A self-locating Wolfram catheter is on the market. It is not clarified if this results in a benefit.The primary aim of this study was to perform a randomized investigation to clarify if the use of a self-locating peritoneal dialysis (PD) catheter would result in different flow problems than a straight Tenckhoff catheter.A total of 61 insertions were made who were randomized and received either a straight Tenckhoff (n = 32) or a self-locating Wolfram catheter (n = 29).A previously described operation technique allowed immediate postoperative start of dialysis. Seven straight Tenckhoff catheters had to be changed into self-locating catheters, and none vice versa, due to flow problems (P = 0.011). An early leakage resulted in temporarily postponed PD in 4 patients.This study showed that using the present operation technique the self-locating PD-catheter causes fewer obstruction episodes than a straight Tenckhoff catheter. This facilitates immediate postoperative start of PD.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.