Abstract

Objective: To assess the influence of the blood content of the tissue and the hemoglobin content of the blood upon pulse oximetry at oxygen saturation ranges that commonly occur in the fetus during birth. Method: A new in-vitro model, which differs conceptually from models used by other authors, was developed to simulate the tissue. Our sensor measured absorption changes in red (640 nm) and infrared (880 nm) light as the blood concentration in the optical chamber changed in analogy to arterial pulsations. Results: Blood content of the tissue and hemoglobin content of the blood both affect the pulse oximetric measurement adversely, especially at low oxygen saturations. Our simulations indicate, for example, that at a true oxygen saturation (Sao2) of 30%, a blood content change from 10 to 6 vol% would cause an underestimation of Sao2 by more than 16% points. Conclusion: Pulse oximetry at low saturations should take the blood content of the tissue and hemoglobin content of the blood into account in order to provide more accurate results.© 1997 Elsevier Science Ireland Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.