Abstract
A commercially produced domiciliary fetal monitoring (DFM) system was assessed in four centres in the UK (Bristol, Glasgow, London and Nottingham) chosen to allow for comprehensive assessment in various settings in many different women. Overall, 825 recordings were made from 368 women (2.24 per woman). There were 56 unsuccessful attempts at transmission (6.8%), most were due to problems with telephone equipment. The system worked most efficiently when a dedicated direct line was used. The data transmission time varied between 40 and 60 s. The median telephone time (including data transmission and conversation) with a dedicated direct line was 7 min. Mean acceptance times from the four centres were between 70 and 80%. All recordings with acceptance times of 40% or more were interpretable. Ten recordings were abnormal. The women and mid-wives were equally proficient at using the DFM system. The DFM system represents an important addition to current methods of fetal assessment. Specific guidelines are outlined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.