Abstract
Prosthetic femoropopliteal grafts attract platelet thrombus and often occlude in the months following implantation. Hence a technique to measure the in vivo thrombogenicity of different vascular substitutes in patients has been developed. In 28 patients the rate of accumulation of 111-indium-labelled platelets in saphenous vein, polytetrafluoroethylene (PTFE) and double velour Dacron femoropopliteal grafts was studied. One week postoperatively autologous 111In-labelled platelets were injected and the isotope emissions over the graft and contralateral leg counted. Graft thrombogenicity was calculated as the daily rise over 7 days in the ratio of counts over the graft to those over the contralateral leg. Mean thrombogenicity index (+/- s.e.mean) was greatest in the Dacron grafts at 0.24 +/- 0.03 which compared with 0.13 +/- 0.02 in PTFE (P less than 0.05). Saphenous vein was markedly less thrombogenic than either prosthetic graft with a mean index of only 0.025 +/- 0.01 (P less than 0.01). Platelet survival failed to differentiate the three graft materials, demonstrating that the direct measurement of 111In platelet deposition was a more sensitive technique.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.