Abstract

Objective To compare radial and femoral artery perfusion pressure during initiation and various stages of cardiopulmonary bypass (CPB).Design Prospective study.Setting The cardiac center of a tertiary referral teaching institute.Participants Sixty consecutive patients of all ages undergoing a variety of cardiac operations.Interventions Radial and femoral arterial pressures were measured in all patients on the same transducer, from the beginning to end of CPB.Measurements and Main Results Mean perfusion pressures on CPB measured at the femoral artery at 1, 5, 10, and 15 minutes of CPB were 38.4 ± 3.6, 46.2 ± 3.1, 49.7 ± 3.9, and 52.8 ± 4.1 mmHg and were significantly greater than the corresponding radial artery pressures (29.9 ± 4.1, 35.3 ± 6.1, 40.9 ± 4.8, and 41.8 ± 5.3 mmHg) (p < 0.001). At 30 minutes and 60 minutes of CPB, femoral artery pressures are higher (60.3 ± 8.8 mmHg and 66.4 ± 8.2 mmHg) compared with radial artery pressures (54.7 ± 6.9 mmHg and 59.6 ± 6.1 mmHg), but the difference is less significant (p < 0.05). On conclusion of CPB, mean femoral artery pressures (70.9 ± 6.7 mmHg) are greater than mean radial artery pressures (67.6 ± 8.1 mmHg) (NS).Conclusions Although radial artery pressures are more commonly monitored during cardiac surgery, femoral artery perfusion pressures are more reliable during the initial part of CPB, and routine monitoring of femoral artery pressures may prevent vasoconstrictor use on initiation of CPB.