Abstract

Using Radio Frequency (RF) waves, magnetic resonance imaging (MRI) systems can make detailed images of human organs for accurate diagnostic processes and disease detection. The overhaul of certain electronic components has become necessary to keep up with the progress of MRI systems. Among the components responsible for image quality, the adjustable non-magnetic capacitors, also called trimmers, are of great importance. The purpose of this paper is to suggest a new design of a trimmer using non-magnetic materials, able to hold a high voltage ( >3 kV) and reach a resonance frequency above 100 MHz. Constructional analysis performed on existing commercial trimmers, combined with the electrical Finite Element Method, were carried out to propose a prototype matching the desired specifications. A dielectric shield surrounding the electrodes was sized to ensure voltage withstand. Moreover, the rotor and the worm-and-gear set were joined together to combat the vibrations generated by the MRI system. The present study includes the modelling step of the component by numerical simulation, the manufacturing, the electrical performance characterization, and the mechanical solution for tuning capacitance. The final prototype combines electrical performance corresponding to the required specifications in particular a high breakdown voltage (39.7 kV) and a resonance frequency (130 MHz).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.