Abstract
The bone-conduction implant (BCI) can use a higher gain setting without having feedback problems compared with a percutaneous bone-conduction device (PBCD). The conventional PBCD, today, is a common treatment for patients with conductive hearing loss and single-sided deafness. However, there are minor drawbacks reported related to the percutaneous implant and specifically poor high-frequency gain. The BCI system is designed as an alternative to the percutaneous system because it leaves the skin intact and is less prone to fall into feedback oscillations, thus allowing more high-frequency gain. Loop gains of the Baha Classic 300 and the BCI were measured in the frequency range of 100 to 10,000 Hz attached to a Skull simulator and a dry cranium. The Baha and the BCI positions were investigated. The devices were adjusted to full-on gain. It was found that the gain headroom using the BCI was generally 0 to 10 dB better at higher frequencies than using the Baha for a given mechanical output. More specifically, if the mechanical output of the devices were normalized at the cochlear level the improvement in gain headroom with the BCI versus the Baha were in the range of 10 to 30 dB. Using a BCI, significantly higher gain setting can be used without feedback problems as compared with using a PBCD.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.