Abstract

A performance comparison of a CdTe and a CdZnTe detector when exposed to uranium samples of various isotopic enrichments has been performed. These high-resolution detectors can assist in the rapid determination of uranium isotopic content of illicit material. Spectra were recorded from these “room temperature' semiconductor detectors with a portable multi-channel analyzer, both in the laboratory and in a field environment. Both detectors were operated below ambient temperature using the vendor supplied thermoelectric coolers. Both detectors had nominally the same active volume (18 mm 3 for the CdZnTe and 25 mm 3 for the CdTe detector) and resolution (i.e., <1.0 keV resolution FWHM at 59.5 keV). Spectra of samples of known isotopic content were recorded at fixed geometries. An evaluation of potential signature g-rays for the detection of enriched uranium was completed. Operational advantages and disadvantages of each detector are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.