Abstract

In this study, we developed a film-type perovskite dosimeter that can be used as a radiotherapy quality assurance (QA) dosimeter while replacing the existing semiconductor dosimeter. Dosimeter performance was evaluated by reproducibility, linearity, dose rate dependence, and percentage depth dose (PDD). In terms of reproducibility, the relative standard deviation (RSD) at 6 MV and 15 MV energy were 1.058% and 1.003%, respectively. As a result of linearity, a coefficient of determination of 0.9999 was presented at 6 MV and 15 MV energies. In terms of dose rate dependence evaluation, the RSD was 1.242% and 0.628% at 6 MV and 15 MV energies, respectively. As for PDD measurement, the result of matching the Dmax point was obtained. Our analysis confirms that the perovskite film dosimeter made of CsPbBr3 was suitable for use as a QA dosimeter in radiotherapy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.