Abstract
We have developed an ultra-low frequency (ULF) nuclear magnetic resonance (NMR) and magnetic resonance imaging (MRI) system utilizing an HTS-SQUID for an application of contaminant detection in food and drink. In the system, a permanent magnet of 1.1T was used to pre-polarize protons in a water sample. We measured NMR signals from water samples with or without various contaminants, such as stainless steel (SUS304), aluminum, and glass balls using the system. In the case that the contaminant was the SUS304 ball, the NMR signal intensity was reduced compared to that from the sample without the contaminant due to the remnant field of the contaminant. One-dimensional (1D) MRIs of the samples were also acquired to detect non-magnetic contaminants. In the 1D MRIs, changes of the MRI spectra were detected, corresponding to positions of the contaminants. These results show that the feasibility of the system to detect various contaminants in foods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.