Abstract

We report on a feasibility study on precise determination of mass-specific activity of low-energy emitting radioisotopes. Conventional methods of activity measurement suffer from source self-absorption and a strong decrease in detection efficiency for low-energy electrons and photons. We propose a new method based on metallic magnetic microcalorimeters with the source embedded in the detector target in a 4π geometry. First results with a 55Fe source show that electrons and photons are detected with a detection efficiency close to unity and with little loss of energy for electrons. The aim of this study is to provide standards of activity with very low uncertainties in the framework of radiation metrology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.