Abstract

The standard of care for neuraxial procedures is landmark palpation to determine the site of placement. This research study aimed to test the primary feasibility of VerTouch™, a force-sensing resistor device, to identify landmarks for the initiation of neuraxial procedures. Patients were recruited at the time of labour epidural, or when consenting for spinal anaesthesia for caesarean delivery at a single centre. The clinical team used the device to create a pressure map image of the bony spine. If they felt confident, they used the device guide to make a mark on the patient’s back. If the mark was used, total insertions and redirections, combined as passes, of the needle during the procedure were counted for secondary outcome analysis. A total of 101 parturients were recruited, and the provider felt confident making a mark based on the imaging in 96.9% of cases. Device success (completion using the mark and ≤4 total passes) occurred in 91.4% of cases. This feasibility study showed that the primary outcome, the provider using the device to make a mark, was successful. In addition, based on comparing passes with historical data, the Vertouch™ device showed promise for future use to minimise needle manipulation in neuraxial procedures.