Abstract

ObjectivesTo examine the feasibility of using the MyotonPRO digital palpation device in measuring the transverse stiffness of tendon tissue.DesignExperimental study.MethodsThe MyotonPRO was used to measure the stiffness and related properties of ballistics gel in comparison with an external materials testing system (PCB electronics). The device was then used to measure the same properties of avian Achilles tendons before and after the removal of the overlying skin and subcutaneous tissue. Next, the test-retest reliability of the Achilles and patellar tendons was determined in humans. Finally, the stiffness of the Achilles tendon was measured before and after competitive running races of varying distances (10, 21 and 42 km, total number of athletes analyzed = 66).ResultsThe MyotonPRO demonstrated a high degree of consistency when testing ballistics gel with known viscoelastic properties. The presence of skin overlying the avian Achilles tendon had a statistically significant impact on stiffness (p<0.01) although this impact was of very small absolute magnitude (with skin; 728 Nm ±17 Nm, without skin; Nm 704 Nm ±7 Nm). In healthy adults of normal body mass index (BMI), the reliability of stiffness values was excellent both for the patellar tendon (ICC, 0.96) and the Achilles tendon (ICC,0.96). In the the field study, men had stiffer tendons than women (p<0.05), and the stiffness of the Achilles tendon tended to increase following running (p = 0.052).ConclusionsThe MyotonPRO can reliably determine the transverse mechanical properties of tendon tissue. The measured values are influenced by the presence of overlying skin, however this does not appear to compromise the ability of the device to record physiologically and clinically relevant measurements.

Highlights

  • The MyotonPRO demonstrated a high degree of consistency when testing ballistics gel with known viscoelastic properties

  • The presence of skin overlying the avian Achilles tendon had a statistically significant impact on stiffness (p

  • The MyotonPRO (Myoton AS, Tallinn, Estonia) is a handheld, digital palpation device which has been used to measure the mechanical properties of muscles and other soft tissue [1]

Read more

Summary

Introduction

The MyotonPRO (Myoton AS, Tallinn, Estonia) is a handheld, digital palpation device which has been used to measure the mechanical properties of muscles and other soft tissue [1]. Using such a device, characterization of the biomechanical properties of the musculoskeletal system has the potential to help identify and diagnose abnormalities in skeletal tissues on-site and in the field without the need of other highly specialized equipment. The tip of the probe is subjected to a constant preload to maintain contact and co-oscillation as the tissue vibrates underneath the skin. An accelerometer linked to the probe generates an acceleration vs. time relation from which various biomechanical characteristics, such as tissue stiffness, can be calculated.[4]

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call