Abstract
We examined the feasibility of capillary liquid chromatography/microchip atmospheric pressure photoionization tandem mass spectrometry (capLC/microAPPI-MS/MS) for the analysis of anabolic steroids in human urine. The urine samples were pretreated by enzymatic hydrolysis (with beta-glucuronidase from Helix pomatia), and the compounds were liquid-liquid extracted with diethyl ether. After separation the compounds were vaporized by microchip APPI, photoionized by a 10 eV krypton discharge lamp, and detected by selected reaction monitoring. The capLC/microAPPI-MS/MS method showed good sensitivity with detection limits at the level of 1.0 ng mL(-1), good linearity with correlation coefficients between 0.9954 and 0.9990, and good repeatability with relative standard deviations below 10%. These results demonstrate that microchip APPI combined with capLC/MS/MS provides a new potential method for analyzing non-polar and neutral compounds in biological samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.