Abstract

This study explored the feasibility of automated characterization of functional mobility via an Instrumented Cane System (ICS) within an older adult sample of cane users. An off-the-shelf offset cane was instrumented with inertial, force, and ultrasound sensors for noninvasive data collection. Eighteen patients from a neurological out-patient rehabilitation clinic and nine independently mobile controls participated in standard clinical evaluations of mobility using the ICS while under the care of an attending physical therapist. Feasibility of the ICS was gauged through two studies. The first demonstrated the capability of the ICS to reliably collect meaningful usage metrics, and the second provided preliminary support for the discriminability of high and low falls risk from system-reported metrics. Specifically, the cane significantly differentiated patients and controls (p < 0.05), and a measure of the variation in rotational velocity was associated with total scores on the Functional Gait Assessment (partial r = 0.61, p < 0.01). These findings may ultimately serve to complement and even extend current clinical assessment practices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.