Abstract

This study evaluated an electromagnetic compatibility method for high-intensity focused ultrasound (HIFU) and magnetic resonance (MR) imaging in an MR-guided focused ultrasound surgery using a conductive wire mesh screen. This screen has a good ultrasound transmission and shielding effectiveness. A hybrid acoustic simulation method was developed to analyze the effects of mesh parameters and the HIFU working frequency on the acoustic field. Experiments were performed to measure both acoustic pressure profile and radiated electromagnetic noise. With the proposed mesh screen, the electromagnetic radiation emission was reduced by 14dB at 128MHz while the acoustic focal intensity was reduced by less than 11% using one screen. This shielding method is easy to implement and requires no additional phase correction method. This method also improves the quality of MR images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.