Abstract

We revisit the doubly-layered sandwich detector configuration for single-shot dual-energy x-ray imaging. In order to understand its proper operation, we investigated the contrast-to-noise performance in terms of the x-ray beam setup using the Monte Carlo methods. Using a pair of active photodiode arrays coupled to phosphor screens, we have built a sandwich detector. For better spectral separation between the projection images obtained from the front and rear detectors during a single x-ray exposure, we inserted a copper sheet between two detectors. We have successfully obtained soft tissue- and bone-enhanced images for a postmortem mouse with the developed sandwich detector using weighted logarithmic subtraction, and the image quality was comparable to those achieved by the conventional kVp-switching technique. Although some problems to be mitigated for the optimal and practical use, for example, the scatter effect and image registration, are still left, the performance of the sandwich detector for single-shot dual-energy x-ray imaging is promising. We expect that the active sandwich detector will provide motion-artifact-free dual-energy images with a reasonable image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.