Abstract

IntroductionHypoxemia and high fractions of inspired oxygen (FiO2) are concerns in critically ill patients. An automated FiO2 controller based on continuous oxygen saturation (SpO2) measurement was tested. Two different SpO2-FiO2 feedback open loops, designed to react differently based on the level of hypoxemia, were compared. The results of the FiO2 controller were also compared with a historical control group.MethodsThe system measures SpO2, compares with a target range (92% to 96%), and proposes in real time FiO2 settings to maintain SpO2 within target. In 20 patients under mechanical ventilation, two different FiO2-SpO2 open loops were applied by a dedicated research nurse during 3 hours, each in random order. The times spent in and outside the target SpO2 values were measured. The results of the automatic controller were then compared with a retrospective control group of 30 ICU patients. SpO2-FiO2 values of the control group were collected over three different periods of 6 hours.ResultsTime in the target range was higher than 95% with the controller. When the 20 patients were separated according to the median PaO2/FiO2 (160(133-176) mm Hg versus 239(201-285)), the loop with the highest slope was slightly better (P = 0.047) for the more-hypoxemic patients. Hyperoxemia and hypoxemia durations were significantly shorter with the controller compared with usual care: SpO2 target range was reached 90% versus 24%, 27% and 32% (P < .001) with the controller, compared with three historical control-group periods.ConclusionA specific FiO2 controller is able to maintain SpO2 reliably within a predefined target range. Two different feedback loops can be used, depending on the initial PaO2/FiO2; with both, the automatic controller showed excellent performance when compared with usual care.Electronic supplementary materialThe online version of this article (doi:10.1186/cc13734) contains supplementary material, which is available to authorized users.

Highlights

  • Hypoxemia and high fractions of inspired oxygen (FiO2) are concerns in critically ill patients

  • When the 20 patients were separated according to the median arterial oxygen partial pressure (PaO2)/fraction of inspired oxygen (FiO2) (160(133-176) mm Hg versus 239(201-285)), the loop with the highest slope was slightly better (P = 0.047) for the more-hypoxemic patients

  • Hyperoxemia and hypoxemia durations were significantly shorter with the controller compared with usual care: SpO2 target range was reached 90% versus 24%, 27% and 32% (P < .001) with the controller, compared with three historical control-group periods

Read more

Summary

Introduction

Hypoxemia and high fractions of inspired oxygen (FiO2) are concerns in critically ill patients. An automated FiO2 controller based on continuous oxygen saturation (SpO2) measurement was tested. Two different SpO2-FiO2 feedback open loops, designed to react differently based on the level of hypoxemia, were compared. The results of the FiO2 controller were compared with a historical control group. It has consequences in case of under- and overdosing. In adult intensive care patients, hypoxemia is a primary preoccupation for all clinicians. The consequences of hyperoxemia are more often neglected because they have been poorly explored. Several clinical observations have suggested that liberal administration of oxygen can be toxic [1,2,3]. Hyperoxia induces the constitution of free oxygen radicals that may cause endothelial cell injury and increases the

Objectives
Methods
Results
Discussion
Conclusion