Abstract

To examine the feasibility of an automated 2-dimensional (2D) to 3- dimensional (3D) image registration system to simplify the navigational challenges faced in complex endovascular aortic procedures. An automated 2D-3D image registration system was used to overlay pre-acquired 3D computed tomography images onto fluoroscopy images taken during endovascular aneurysm repair. Errors between the 3D overlay and digital subtraction angiograms were measured and correlated with aortic neck angulation. A mean discrepancy < or =3 mm was considered clinically acceptable. There was a strong correlation between maximum neck angulation and maximum registration error (Pearson's r = 0.75). Aortas with a maximum neck angulation < or =30 degrees had a mean error of 2.5+/-1.2 mm, whereas aortas with neck angulation >30 degrees had a mean error of 6.2+/-2.5 mm (p<0.0001). The major source of registration errors is aortic deformation caused by the presence of the introducer and endovascular graft. Further work is required if this technology is to be routinely applied to severely angulated aortic anatomy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.