Abstract

The medical community has no independent source of timely information regarding the performance of pacemaker and ICD pulse generators and leads. Accordingly, the authors established an Internet-based registry of pacemaker and ICD pulse generator and lead failures (www.pacerandicregistry.com). During the first year, they found three previously unreported device problems that were promptly communicated to the participants. Of the failures reported, 11% of ICD and 10% of pacemaker pulse generator failures were heralded by signs other than the expected elective replacement indicator (ERI). Average ICD battery longevity was 4.0 +/- 0.7 years, and average dual chamber pacemaker battery longevity was 6.8 +/- 2.6 years. Disrupted insulation accounted for 54% of pacemaker and 29% of ICD lead failures. Compared to pacemaker pulse generator and lead failure, ICD device failures were more likely to cause severe clinical consequences. In conclusion, an Internet-based registry is feasible and capable of providing timely data regarding the signs, causes, and clinical consequences of pacemaker and ICD failures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.