Abstract

To determine whether time-resolved immunofluorescence assay (TRIFA) shares the similar accuracy and specificity with automatic chemiluminescence immunoassay (CMIA) in analyzing HBeAg levels in hepatitis B. A total of 157 serum samples were collected from outpatients with infection of HBV in Xiangya Hospital, Central South University. CMIA and TRIFA were used to analyze HBeAg quantitation and HBeAg/HBeAb qualitative detection, respectively. The linear regression equation for the two methods was Y=0.72779X-4.0551 (r=0.712, P<0.001). Compared with the CMIA, the sensitivity and specificity in detection of HBeAg by TRIFA were 89.89% and 100%, respectively, and the coincidence rate of HBeAg was 94.27% by two assays. Similarly, the sensitivity and specificity in detection of HBeAb by TRIFA were 100% and 95.45%, respectively. The coincidence rate was 97.45% by two assays. TRIFA has similar accuracy, sensitivity, and specificity with CMIA in quantitative detection of HBeAg, and their coincidence rate in detection of HBeAg/HBeAb is high.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.