Abstract
Based on product simulation software, FEA method is used to identify the structural modal of instrumented indentation equipment frame which is the main supporting part of the High-precision instrumented indentation tester developed and realized by our group. From dynamic finite element computations, the multistage natural frequencies of tester frame are derived. The first 6 ranks vibration frequencies are all less than 10Hz, therefore, the loading and unloading process should be kept in quasi static state so as to prevent system resonance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: DEStech Transactions on Computer Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.